Remote control commands
R&S
®
NRPxxA(N)
88
User Manual 1177.6017.02 ─ 08
Parameters:
<resolution>
Range:
1 to 4
*RST:
3
Manual operation:
See
[SENSe<Sensor>:]AVERage:RESet
Deletes all previous measurement results that the averaging filter contains and initial-
izes the averaging filter. The filter length gradually increases from 1 to the set averag-
ing factor. Thus, trends in the measurement result become quickly apparent. Note that
the measurement time required for the averaging filter to settle completely remains
unchanged.
Use this command if:
●
High averaging factor is set.
●
Intermediate values are output as measurement results.
[SENSe<Sensor>:]AVERage:TCONtrol
MOVing
●
Power has significantly decreased since the previous measurement, for example
by several powers of 10.
In this situation, previous measurement results, which are still contained in the averag-
ing filter, strongly affect the settling of the display. As a result, the advantage of detect-
ing trends in the measurement result while the measurement is still in progress is lost.
Example:
AVER:RES
Usage:
Event
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE
<type>
Sets the automatic averaging filter mode.
Parameters:
<type>
RESolution | NSRatio
RESolution
The usual mode for the power sensors.
NSRatio
Predefines the compliance to an exactly defined noise compo-
nent.
*RST:
RESolution
Manual operation:
See
[SENSe<Sensor>:]AVERage:TCONtrol
<mode>
Defines how the measurement results are output. This is called termination control.
Configuring basic measurement parameters