Preparing for use
R&S
®
NRPxxA(N)
17
User Manual 1177.6017.02 ─ 08
See
Chapter 3.7.3, "Using a LAN connection"
3.7.1.1
Simple USB connection
All R&S NRPxxA(N) power sensors can be connected to the USB interface of a com-
puter.
Required equipment
●
R&S NRPxxA(N) power sensor
●
R&S
NRP
‑
ZKU cable
Setup
NRP
3-Path Diode Power Sensor
MHz to GHz, 100 pW to 200 mW (−70 dBm to +23 dBm)
SMART SENSOR TECHNOLOGY
1
2
3
4
5
6
Figure 3-1: Setup with an R&S
NRP
‑
ZKU cable
1 = Signal source
2 = R&S NRPxxA(N) power sensor
3 = Host interface connector
4 = R&S
NRP
‑
ZKU cable
5 = USB connector
6 = Computer with installed VISA driver or R&S
NRP Toolkit
.
1. Connect the R&S
NRP
‑
ZKU cable to the power sensor. See
the host interface of the power sensor"
2. Connect the R&S
NRP
‑
ZKU cable to the computer.
3.
NOTICE!
Incorrectly connecting or disconnecting the power sensor can damage
the power sensor or lead to erroneous results. Ensure that you connect or discon-
nect the power sensor as described in
Chapter 3.4, "Connecting to a DUT"
on page 14.
Connect the power sensor to the signal source.
4. On the computer, start a software application to view the measurement results.
Chapter 5, "Operating concepts"
Connecting to a controlling host