Additional Basics on Remote Control
R&S
®
NGL200/NGM200
164
User Manual 1178.8736.02 ─ 06
Figure A-1: Structure of the status QUEStionable register
CONDition
●
The
CONDition
register queries the actual state of the instrument. If you want to
query the constant voltage or current mode, you have to use the
CONDition
regis-
ter.
The
CONDition
register delivers a "1" (first bit set) in constant current mode (CC) and
a "2" (second bit set) in constant voltage mode (CV).
If the correct channel is selected and the red LED of the channel button lights up (CC
mode), the query of the
CONDition
register must deliver a "1".
Example:
STAT:QUES:ISUM1:COND?
EVENt
●
The
EVENt
status register is set (1) until it is queried. After reading (query), the
EVENt
status register is set to zero.
Status Reporting System