Instrument Functions
R&S
®
NGL200
62
User Manual 1178.8736.02 ─ 02.01
Figure 6-23: Target folder dialog
● "USB1A": Target folder is set to
/USB1/NGM202/logging
.
● "int": Target folder is set to
/int/logging
.
7. Set the required target folder.
8. Select "Set" to confirm the selection.
9. Select "Mode" to set logging duration.
● "Duration": Time taken for data logging with duration and time interval setting.
● "Span": Time taken for data logging with start time, time interval and duration
setting
● "Unlimited": Data logging with time interval setting. The data logging continuous
until function is deactivated.
● "Count": Data logging with number of counts and time interval setting
10. Depending on the mode selected, configure the required settings for the logging
duration.
11. Set the "CSV Settings".
12. Only applicable with R&S NGL-K103, Digital Output option, see
Activated the "Triggered" if logging of data is to be excuted under triggered condi-
tions.
If activated, the R&S
NGL200 will execute the logging of data when a trigger is
detected.
13. Select "Enabled" to activate the file logging function.
The R&S
NGL200 starts the data logging and displays the "data logging" icon at
the device status bar information.
Note: The data logging dialog is not configurable if logging is enabled.
6.7.1
CSV Settings
Data Logger