Remote Control Commands
R&S
®
NGL200
112
User Manual 1178.8736.02 ─ 02.01
7.6
Measurement Commands
The
MEASure
subsystem provides commands to query the voltage and current values
of a channel.
MEASure[:SCALar]:CURRent[:DC]?
.................................................................................112
......................................................................................... 112
................................................................................. 112
................................................................................. 112
........................................................................................... 113
MEASure[:SCALar][:VOLTage][:DC]?
............................................................................... 113
MEASure[:SCALar]:CURRent[:DC]?
Queries the currently measured current of the selected channel.
Example:
MEAS:CURR?
-> 1.000E +00
Usage:
Query only
MEASure[:SCALar]:ENERgy?
Queries the measured the current released energy value of the previous selected
channel. Please notice that the energy meter has to be activated to measure the cur-
rent released energy value.
Example:
MEAS:ENER?
-> 5.382E+00 (value in Wh)
Usage:
Query only
MEASure[:SCALar]:ENERgy:RESet
Resets the energy counter for the selected channel.
Usage:
Event
MEASure[:SCALar]:ENERgy:STATe
<state>
MEASure[:SCALar]:ENERgy:STATe?
Activates/Deactivates or Queries the energy counter for the selected channel.
Parameters:
<state>
OFF | ON | 0 | 1
ON |1
Activates the energy counter.
OFF | 0
Deactivates the energy counter.
Measurement Commands