Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
117
User Manual 1177.5862.02 ─ 09
Suffix:
<n>
.
irrelevant
Example:
INIT:REFR
Updates the IQ measurement results.
Manual operation:
See
[SENSe:]MSRA:CAPTure:OFFSet
<Offset>
This setting is only available for secondary applications in MSRA mode, not for the
MSRA primary application. It has a similar effect as the trigger offset in other measure-
ments.
Parameters:
<Offset>
This parameter defines the time offset between the capture buf-
fer start and the start of the extracted secondary application
data. The offset must be a positive value, as the secondary
application can only analyze data that is contained in the cap-
ture buffer.
Range:
0 to <Record length>
*RST:
0
Default unit: S
Manual operation:
See
Commands specific to MSRA secondary applications