Remote Commands
R&S
®
FSW
956
User Manual 1173.9411.02 ─ 43
[SENSe:]LIST:RANGe<ri>:DELete
This command removes a range from the sweep list.
Note that
●
you cannot delete the reference range
●
a minimum of three ranges is mandatory.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Example:
LIST:RANG2:DEL
[SENSe:]LIST:RANGe<ri>:DETector
<Detector>
This command selects the detector for a spurious emission measurement range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Detector>
APEak
Autopeak
NEGative
minimum peak detector
POSitive
peak detector
SAMPle
sample detector
RMS
RMS detector
AVERage
average detector
*RST:
RMS
Example:
LIST:RANG2:DET AVER
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:FILTer:TYPE
<FilterType>
This command selects the filter type for a spurious emission measurement range.
The EMI-specific filter types are available if the EMI (R&S
FSW-K54) measurement
option is installed, even if EMI measurement is not active. For details see
ter 7.13.3.1, "Resolution Bandwidth and Filter Types"
Configuring and Performing Measurements
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...