Remote Commands
R&S
®
FSW
955
User Manual 1173.9411.02 ─ 43
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:BANDwidth:VIDeo
<VBW>
This command defines the video bandwidth for a spurious emission measurement
range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<VBW>
Video bandwidth.
Refer to the data sheet for available video bandwidths.
Default unit: Hz
Example:
LIST:RANG2:BAND:VID 3KHZ
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:BREak
<State>
This command controls the sweep for all ranges.
Suffix:
<ri>
.
irrelevant
Parameters:
<State>
ON | 1
The R&S
FSW stops after measuring one range, and the status
bit number 10 in the
STAT:OPER
register is set.
(See
To continue with the next range, use
.
OFF | 0
The R&S
FSW sweeps all ranges in one go.
*RST:
0
Example:
LIST:RANG2:BRE ON
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:COUNt?
This command queries the number of ranges in the sweep list.
Suffix:
<ri>
.
irrelevant
Return values:
<Ranges>
Number of ranges in the sweep list.
Example:
LIST:RANG:COUN?
Usage:
Query only