Remote Commands
R&S
®
FSW
923
User Manual 1173.9411.02 ─ 43
<ri>
1..n
Selects the measurement range.
Parameters:
<VBW>
Video bandwidth.
Refer to the data sheet for available video bandwidths.
*RST:
10.0 MHz
Default unit: Hz
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:COUNt?
This command queries the number of ranges in the sweep list.
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
<ri>
irrelevant
Return values:
<Ranges>
Number of ranges in the sweep list.
Usage:
Query only
[SENSe:]ESPectrum<sb>:RANGe<ri>:DELete
This command removes a range from the sweep list.
Note that
●
you cannot delete the reference range
●
a minimum of three ranges is mandatory.
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
<ri>
Selects the measurement range.
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:FILTer:TYPE
<FilterType>
This command selects the filter type for an SEM range.
In case of high speed measurements, the filter has to be identical for all ranges.
The EMI-specific filter types are available if the EMI (R&S
FSW-K54) measurement
option is installed, even if EMI measurement is not active. For details see
ter 7.13.3.1, "Resolution Bandwidth and Filter Types"
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement