Remote Commands
R&S
®
FSW
830
User Manual 1173.9411.02 ─ 43
You can start a new sequence any time using
Usage:
Event
Manual operation:
See
INITiate:SEQuencer:IMMediate
This command starts a new sequence of measurements by the Sequencer.
Its effect is similar to the
measurement.
Before this command can be executed, the Sequencer must be activated (see
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single sequence mode so each active measurement will be
performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
Manual operation:
See
INITiate:SEQuencer:MODE
<Mode>
Defines the capture mode for the entire measurement sequence and all measurement
groups and channels it contains.
Note:
In order to synchronize to the end of a measurement sequence using *OPC,
*OPC? or *WAI you must use
SINGle
Sequence mode.
Parameters:
<Mode>
SINGle
Each measurement group is started one after the other in the
order of definition. All measurement channels in a group are
started simultaneously and performed once. After
all
measure-
ments are completed, the next group is started. After the last
group, the measurement sequence is finished.
CONTinuous
Each measurement group is started one after the other in the
order of definition. All measurement channels in a group are
started simultaneously and performed once. After
all
measure-
ments are completed, the next group is started. After the last
group, the measurement sequence restarts with the first one and
continues until it is stopped explicitely.
*RST:
CONTinuous
Manual operation:
See
Selecting the Operating Mode and Application