Getting Started
R&S
®
FSW
61
User Manual 1173.9411.02 ─ 43
5.2.2.13
AUX PORT
The 9-pole SUB-D male connector provides control signals for controlling external
devices. The voltage levels are of the TTL type (max. 5
V).
Pin
Signal
Description
1
+5 V / max. 250 mA
Supply voltage for external circuits
2 to 7
I/O
Control lines for user ports (see User manual)
8
GND
Ground
9
READY FOR TRIGGER
Signal indicating that the instrument is ready to
receive a trigger signal (Low active =
0
V)
Short-circuit hazard
Always observe the designated pin assignment. A short-circuit can damage the port.
5.2.2.14
GPIB Interface
The GPIB interface is in compliance with IEEE488 and SCPI. A computer for remote
control can be connected via this interface. To set up the connection, a shielded cable
is recommended. For more details, refer to "Setting Up Remote Control" in the User
Manual.
5.2.2.15
External Generator Control Option (R&S
FSW-B10)
The external generator control option provides an additional GPIB and an "AUX con-
trol" connector.
The GPIB connector can be used to connect the external generator to the R&S
FSW.
The female "AUX control" connector is required for TTL synchronization, if supported
by the generator.
For details on connecting an external generator, see the "External Generator Control"
section of the R&S
FSW User Manual.