Getting Started
R&S
®
FSW
30
User Manual 1173.9411.02 ─ 43
Performing a self-test
You do not have to repeat the self-test every time the instrument is switched on. It is
only necessary when instrument malfunction is suspected.
1. Press the [SETUP] key.
2. Press the "Service" softkey.
3. Switch to the "Selftest" tab in the "Service" dialog box.
4. Select the "Start Selftest" button.
Once the instrument modules have been checked successfully, a message is dis-
played.
5.1.1.7
Checking the Supplied Options
The instrument may be equipped with both hardware and firmware options. To check
whether the installed options correspond to the options indicated on the delivery note,
proceed as follows.
1. Press the [SETUP] key.
2. Press the "System Config" softkey.
3. Switch to the "Ve Options" tab in the "System Configuration" dialog box.
A list with hardware and firmware information is displayed.
4. Check the availability of the hardware options as indicated in the delivery note.
5.1.2
Windows Operating System
The instrument contains the Windows
10 operating system which has been configured
according to the instrument's features and needs. Changes in the system setup are
only required when peripherals like keyboard or a printer are installed or if the network
configuration does not comply with the default settings. After the R&S
FSW is started,
the operating system boots and the instrument firmware is started automatically.
To ensure that the instrument software functions properly, certain rules must be
adhered to concerning the operating system.
Tested software
The drivers and programs used on the instrument under Windows
10 are adapted to
the instrument. Only install update software released by Rohde
&
Schwarz to modify
existing instrument software.
You can install additional software on the instrument, however, additional software can
impair instrument function. Thus, run only programs that Rohde
&
Schwarz has tested
for compatibility with the instrument software.