Getting Started
R&S
®
FSW
28
User Manual 1173.9411.02 ─ 43
Rackmounting
The R&S
FSW can be installed in a rack using a rack adapter kit (order no. see data
sheet). The installation instructions are part of the adapter kit.
Risk of instrument damage due to insufficient airflow in a rack
If you mount several instruments in a rack, you need an efficient ventilation concept to
ensure that the instruments do not overheat. Insufficient airflow for a longer period can
disturb the operation and even cause damage.
5.1.1.4
Connecting the AC Power
In the standard version, the R&S
FSW is equipped with an AC power supply connector.
The R&S
FSW can be used with different AC power voltages and adapts itself auto-
matically to it. Refer to the datasheet for the requirements of voltage and frequency.
The AC power connector is located on the rear panel of the instrument.
For details on the connector, refer to
Chapter 5.2.2.2, "AC Power Supply Connection
► Connect the R&S
FSW to the AC power supply using the supplied power cable.
Since the instrument is assembled in line with the specifications for safety class
EN61010, it must only be connected to an outlet that has a ground contact.
5.1.1.5
Switching the Instrument On and Off
Switching on the instrument
► Press the AC power switch on the rear panel to position "I".
The instrument is supplied with AC power. After booting, the instrument is ready for
operation. A green LED above the [POWER] key indicates this.
An orange LED indicates the instrument is in standby mode
Warm-up time for OCXO
When the instrument is switched on, the OCXO requires an extended warm-up time
(see data sheet).
Switching off the instrument
1. Press the [POWER] key on the front panel.
The R&S
FSW switches to standby mode.
Preparing for Use