Remote Commands
R&S
®
FSW
1230
User Manual 1173.9411.02 ─ 43
Example:
MMEM:CAT? 'C:\R_S\INSTR\USER\SPOOL?.PNG'
Returns all files in
C:\R_S\INSTR\USER
whose names start
with SPOOL, have 6 letters and the extension
.PNG
, e.g.:
SPOOL1.PNG,SPOOL2.PNG,SPOOL3.PNG
Example:
MMEM:CAT? 'C:\R_S\INSTR\USER\SPOOL6.PNG'
Query whether the file '
SPOOL6.PNG
' also exists in the directory;
Result:
-256,"File name not found;:MMEMory:CATalog?
'C:\R_S\INSTR\USER\SPOOL6.PNG'
Manual operation:
See
"Selecting Storage Location - Drive/ Path/ Files"
MMEMory:CATalog:LONG
<arg0>
This command returns the contents of a particular directory with additional information
about the files.
Parameters:
<arg0>
String containing the path and directory.
If you leave out the path, the command returns the contents of
the directory selected with
on page 1230.
The path may be relative or absolute. Using wildcards ('*') is
possible to query a certain type of files only.
MMEMory:CDIRectory
<Directory>
This command changes the current directory.
Parameters:
<Directory>
String containing the path to another directory.
The path may be relative or absolute.
MMEMory:COMMent
<arg0>
This command defines a comment for the stored settings.
Parameters:
<arg0>
String containing the comment.
Example:
MMEMory:COMMent "ACP measurement with Standard
Tetra from 23.05."
MMEMory::MMEMory:STORe1:STATe 1, "ACP_T"
As a result, in the selection list for recall settings, the comment
"ACP measurement with Standard Tetra from
23.05."
is added to the ACP entry.
Manual operation:
See
Managing Settings and Results