CALCulate<n>:LIMit<li>:UPPer:THReshold
<Limit>
This command defines an absolute limit for limit lines with a relative scale.
The R&S
FSW uses the threshold for the limit check, if the limit line violates the thresh-
old.
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<Limit>
Numeric value.
The unit depends on
on page 1218.
*RST:
-200
Default unit: dBm
Manual operation:
See
Managing Limit Lines
Useful commands for managing limit lines described in the R&S
FSW User Manual:
●
MMEM:SEL[:ITEM]:LIN:ALL
●
MMEM:STOR:TYPE
●
MMEM:LOAD:TYPE
Remote commands exclusive to managing limit lines:
CALCulate<n>:LIMit<li>:ACTive?
...................................................................................1221
......................................................................................1222
.................................................................................... 1222
......................................................................................1222
CALCulate<n>:LIMit<li>:TRACe<t>:CHECk
.................................................................... 1223
............................................................................................ 1223
.......................................................................................... 1223
CALCulate<n>:LIMit<li>:ACTive?
This command queries the names of
all
active limit lines.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
Return values:
<LimitLines>
String containing the names of all active limit lines in alphabeti-
cal order.
Example:
CALC:LIM:ACT?
Queries the names of all active limit lines.
Usage:
Query only
Analyzing Measurements (Basics)