Remote Commands
R&S
®
FSW
1095
User Manual 1173.9411.02 ─ 43
–
"active modular"
Example:
//Query probe type
PROB3:SET:TYPE?
Usage:
Query only
Manual operation:
See
14.7.6.4
External Generator Control
External generator control commands are available if the R&S
FSW External Genera-
tor Control option (R&S
FSW-B10) is installed.
For each measurement channel one external generator can be configured. To switch
between different configurations define multiple measurement channels.
For more information on external generator control see
................................................................................1095
........................................................................................1099
...............................................................................................1101
Programming Example for External Generator Control
.......................................1104
Measurement Configuration
The following commands are required to activate external generator control and to con-
figure a calibration measurement with an external tracking generator.
SOURce<si>:EXTernal<gen>:FREQuency
......................................................................1095
SOURce<si>:EXTernal<gen>:FREQuency:COUPling[:STATe]
.......................................... 1096
SOURce<si>:EXTernal<gen>:FREQuency[:FACTor]:DENominator
....................................1096
SOURce<si>:EXTernal<gen>:FREQuency[:FACTor]:NUMerator
....................................... 1097
SOURce<si>:EXTernal<gen>:FREQuency:OFFSet
......................................................... 1097
SOURce<si>:EXTernal<gen>:POWer[:LEVel]
................................................................. 1098
SOURce<si>:EXTernal<gen>[:STATe]
............................................................................ 1098
SOURce<si>:POWer[:LEVel][:IMMediate]:OFFSet
...........................................................1099
SOURce<si>:EXTernal<gen>:FREQuency
<Frequency>
This command defines a fixed source frequency for the external generator.
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<Frequency>
Source frequency of the external generator.
*RST:
1100050000
Default unit: HZ
Example:
//Define frequency of the generator
SOUR:EXT:FREQ 10MHz