Remote Commands
R&S
®
FSW
1086
User Manual 1173.9411.02 ─ 43
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:MIX 'FS_Z60'
Manual operation:
See
[SENSe:]CORRection:CVL:PORTs
<PortType>
This command defines the mixer type in the conversion loss table. This setting is
checked against the current mixer setting before the table can be assigned to the
range.
Before this command can be performed, the conversion loss table must be selected
(see
This command is only available with option B21 (External Mixer) installed.
Parameters:
<PortType>
2 | 3
*RST:
2
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:PORT 3
Manual operation:
See
[SENSe:]CORRection:CVL:SELect
<FileName>
This command selects the conversion loss table with the specified file name. If
<file_name> is not available, a new conversion loss table is created.
This command is only available with option B21 (External Mixer) installed.
Parameters:
<FileName>
String containing the path and name of the file.
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Manual operation:
See
See
See
[SENSe:]CORRection:CVL:SNUMber
<SerialNo>
This command defines the serial number of the mixer for which the conversion loss
table is to be used. This setting is checked against the current mixer setting before the
table can be assigned to the range.
Before this command can be performed, the conversion loss table must be selected
(see
This command is only available with option B21 (External Mixer) installed.
Parameters:
<SerialNo>
Serial number with a maximum of 16 characters