Remote Commands
R&S
®
FSW
1064
User Manual 1173.9411.02 ─ 43
//0.005
SWE:EGAT:CONTinuous:PLENgth 4MS
SWE:EGAT:CONTinuous:PLENgth?
//0.004
SWE:EGAT:CONTinuous:PCOunt?
//100
SWE:EGAT:CONTinuous:PCOunt 80
SWE:EGAT:CONTinuous:PCOunt?
//80
//--------------Performing the Measurement-----
INIT:CONT OFF
INIT
// Sweep duration is less than 1 second
14.7.5
Adjusting Settings Automatically
The commands required to adjust settings automatically in a remote environment are
described here.
The tasks for manual operation are described in
Chapter 8.7, "Adjusting Settings Auto-
MSRA operating mode
In MSRA operating mode, settings related to data acquisition (measurement time, hys-
teresis) can only be adjusted automatically in the MSRA Master, not in the MSRA
applications.
[SENSe:]ADJust:CONFigure:DURation
.......................................................................... 1065
[SENSe:]ADJust:CONFigure:DURation:MODE
................................................................1065
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer
............................................................ 1065
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer
............................................................. 1066
[SENSe:]ADJust:CONFigure:TRIGger
............................................................................1066
........................................................................................1066
[SENSe:]ADJust:ALL
This command initiates a measurement to determine and set the ideal settings for the
current task automatically (only once for the current measurement).
This includes:
●
Center frequency
●
Reference level
Example:
ADJ:ALL
Setting Basic Measurement Parameters