Remote Commands
R&S
®
FSW
1025
User Manual 1173.9411.02 ─ 43
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:X:SPACing
<Scale>
This command selects the scaling of the x-axis.
Suffix:
<n>
.
<w>
subwindow
<t>
Parameters:
<Scale>
LOGarithmic
Logarithmic scaling.
LINear
Linear scaling.
*RST:
LINear
Example:
DISP:TRAC:X:SPAC LOG
Manual operation:
See
[SENSe:]FREQuency:CENTer
<Frequency>
This command defines the center frequency.
Parameters:
<Frequency>
The allowed range and f
max
is specified in the data sheet.
UP
Increases the center frequency by the step defined using the
DOWN
Decreases the center frequency by the step defined using the
*RST:
fmax/2
Default unit: Hz
Example:
FREQ:CENT 100 MHz
FREQ:CENT:STEP 10 MHz
FREQ:CENT UP
Sets the center frequency to 110 MHz.
Manual operation:
See
See
See
[SENSe:]FREQuency:CENTer:STEP
<StepSize>
This command defines the center frequency step size.
You can increase or decrease the center frequency quickly in fixed steps using the
SENS:FREQ UP
AND
SENS:FREQ DOWN
Setting Basic Measurement Parameters
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...