Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
205
User Manual 1173.9240.02 ─ 23
SENS:MIX:LOSS:HIGH 30dB
//Query the possible range
SENS:MIX:FREQ:STAR?
//Result: 47480000000 (47.48 GHz)
SENS:MIX:FREQ:STOP?
//Result: 138020000000 (138.02 GHz)
//--------------Performing the Measurement-----
//Select single sweep mode.
INIT:CONT OFF
//Initiate a basic frequency sweep and wait until the sweep has finished.
INIT;*WAI
//---------------Retrieving Results-------------
//Return the trace data (default screen configuration)
TRAC:DATA? TRACe1
11.4.2.3
Configuring the 2 GHz Bandwidth Extension (R&S
FSW-B2000)
The following commands are required to use the optional 2
GHz bandwidth extension
(R&S
FSW-B2000).
See also the command for configuring triggers while using the optional 2
GHz band-
width extension (R&S
FSW-B2000):
●
TRIGger[:SEQuence]:OSCilloscope:COUPling
Remote commands exclusive to configuring the 2
GHz bandwidth extension:
......................................................................................... 205
SYSTem:COMMunicate:RDEVice:OSCilloscope[:STATe]
.................................................. 206
SYSTem:COMMunicate:RDEVice:OSCilloscope:ALIGnment:STEP[:STATe]?
SYSTem:COMMunicate:RDEVice:OSCilloscope:ALIGnment:DATE?
.................................. 207
SYSTem:COMMunicate:RDEVice:OSCilloscope:IDN?
.......................................................207
SYSTem:COMMunicate:RDEVice:OSCilloscope:LEDState?
...............................................207
SYSTem:COMMunicate:RDEVice:OSCilloscope:SRATe
....................................................208
SYSTem:COMMunicate:RDEVice:OSCilloscope:TCPip
..................................................... 208
SYSTem:COMMunicate:RDEVice:OSCilloscope:VDEVice?
............................................... 208
SYSTem:COMMunicate:RDEVice:OSCilloscope:VFIRmware?
........................................... 209
TRIGger[:SEQuence]:OSCilloscope:COUPling
..................................................................209
EXPort:WAVeform:DISPlayoff
<FastExport>
Enables or disables the display update on the oscilloscope during data acquisition with
the
optional 2
GHz bandwidth extension (R&S
FSW-B2000)
.
As soon as the R&S
FSW-B2000 is activated (see
play on the oscilloscope is turned off to improve performance during data export. As
soon as the R&S
FSW closes the connection to the oscilloscope, the display is reacti-
vated and the oscilloscope can be operated as usual. However, if the LAN connection
is lost for any reason, the display of the oscilloscope remains deactivated. Use this
command to re-activate it.
Configuring the Measurement