Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
932
User Manual 1178.8520.02 ─ 01
<gen>
Parameters:
<Value>
<numeric value>
*RST:
1
Example:
//Define multiplication factor of 4/3; the transmit frequency of the
generator is 4/3 times the analyzer frequency
SOUR:EXT:FREQ:NUM 4
SOUR:EXT:FREQ:DEN 3
Manual operation:
See
"(Automatic) Source Frequency (Numerator/Denominator/
SOURce<si>:EXTernal<gen>:FREQuency[:FACTor]:NUMerator
<Value>
This command defines the numerator of the factor with which the analyzer frequency is
multiplied in order to obtain the transmit frequency of the selected generator.
Select the multiplication factor such that the frequency range of the generator is not
exceeded if the following formula is applied to the start and stop frequency of the ana-
lyzer:
Offset
Analyzer
Generator
F
ator
Deno
Numerator
F
F
min
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<Value>
<numeric value>
*RST:
1
Example:
//Define multiplication factor of 4/3; the transmit frequency of the
generator is 4/3 times the analyzer frequency
SOUR:EXT:FREQ:NUM 4
SOUR:EXT:FREQ:DEN 3
Manual operation:
See
"(Automatic) Source Frequency (Numerator/Denominator/
SOURce<si>:EXTernal<gen>:FREQuency:OFFSet
<Offset>
This command defines the frequency offset of the generator with reference to the ana-
lyzer frequency.
Select the offset such that the frequency range of the generator is not exceeded if the
following formula is applied to the start and stop frequency of the analyzer:
Offset
Analyzer
Generator
F
ator
Deno
Numerator
F
F
min
Setting Basic Measurement Parameters