Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
868
User Manual 1178.8520.02 ─ 01
Note that using the command as a query initiates the measurement and returns the
results if all frequencies have been measured. For more information on querying the
results see
Parameters:
<Frequency>
Defines the frequency. Each frequency corresponds to one list
entry.
Range:
0 to Fmax
Default unit: Hz
<RefLevel>
Defines the reference level for a list entry.
Range:
-130 to 30
Increment: 0.01
Default unit: dBm
<RFAttenuation>
Defines the RF attenuation for a list entry.
Range:
0 to 70
Increment: 1
Default unit: dB
<FilterType>
Selects the filter type for a list entry. For more information see
[SENSe:]BANDwidth[:RESolution]:TYPE
<RBW>
Defines the resolution bandwidth for a list entry.
<VBW>
Defines the measurement time for a list entry.
<MeasTime>
Defines the measurement time for a list entry.
Range:
1 µs to 16000 s
Default unit: s
<TriggerLevel>
The trigger level must be 0.
<PowerLevel>
Power level for each frequency included in the measurement.
The command returns up to 3 power levels for each frequency,
depending on the number of evaluation modes you have turned
on with
The result is a list of floating point values separated by commas.
The unit depends on
[SENSe:]LIST:POWer:SET
<PeakPower>, <RMSPower>, <AVGPower>,
<TriggerSource>, <TriggerSlope>, <TriggerOffset>, <GateLength>
This command defines global List Evaluation parameters.
These parameters are valid for every frequency you want to measure.
The state of the first three parameters (<PeakPower>, <RMSPower> and <AVG-
Power>) define the number of results for each frequency in the list.
Note that you have to set the trigger level
after
sending this command.
Configuring and Performing Measurements