Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
864
User Manual 1178.8520.02 ─ 01
14.5.12
Measuring the AM Modulation Depth
All remote control commands specific to AM modulation depth measurements are
described here.
Configuring and Performing the Measurement
..................................................... 864
Example: Measuring the AM Modulation Depth
....................................................865
14.5.12.1
Configuring and Performing the Measurement
The following commands control the measurement.
Useful commands for AM modulation depth described elsewhere
●
●
CALCulate<n>:DELTamarker<m>:X:RELative?
●
Remote commands exclusive to AM modulation depth measurements
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth[:STATe]
................................................... 864
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth:SEARchsignal ONCE
............................... 864
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth:RESult<t>?
..............................................865
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth[:STATe]
<State>
This command turns the AM Modulation Depth measurement on and off.
To work correctly, the measurement requires an AM modulated signal.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth:SEARchsignal ONCE
This command initiates a search for the signals required for the AM depth measure-
ment.
Note that the command does not perform a new measurement, but looks for the sig-
nals on the current trace.
Suffix:
<n>
.
<m>
Configuring and Performing Measurements