General Instrument Setup
R&S
®
FSVA3000/ R&S
®
FSV3000
568
User Manual 1178.8520.02 ─ 01
Deactivating and Activating the Touchscreen
............................................................ 568
............................................................................................... 568
Deactivating and Activating the Touchscreen
The touchscreen function can be deactivated, e.g. when the instrument is being used
for demonstration purposes and tapping the screen must not provoke an action.
To reactivate the touchscreen, simply press the [Setup] key on the front panel. The
"Display" dialog box is opened automatically and the "Touch Screen" option is set to
"On" .
"Touch On"
Touchscreen function is active for the entire screen.
" Touch Off"
Touchscreen is deactivated for the entire screen.
" Touch Diagram Off"
Touchscreen is deactivated for the diagram area of the screen, but
active for the surrounding softkeys, toolbars and menus.
Remote command:
Display Update Rate
By default, a fast update rate ensures the most recent measurement results on the dis-
play. However, when performance is poor due to slow data transfer (for example during
remote control), it can be helpful to decrease the frequency with which the screen dis-
play is updated.
Set Date and Time
The current date and time on the instrument is set using the standard Windows "Date
and Time Properties" dialog box. Select the "Set Date and Time" button in the "Display"
dialog box, or select the date and time display in the status bar to open the Windows
dialog box.
Date and Time Format
Switches the time and date display on the screen between US and German (DE) for-
mat.
Remote command:
DISPlay[:WINDow<n>]:TIME:FORMat
12.2.1.2
Displayed Items
Access
: [Setup] > "Display" > "Displayed Items"
Several elements on the screen display can be hidden or shown as required, for exam-
ple to enlarge the display area for the measurement results.
Display Settings