Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
271
User Manual 1178.8536.02 ─ 01
Query parameters:
<TraceNumber>
Trace number.
TRACE1 | ... | TRACE6
Example:
TRAC3:X? TRACE1
Returns the x-values for trace 1 in window 3.
Usage:
Query only
8.8.3
Exporting Traces and Data
The following commands are required to export traces and spectrograms.
........................................................................................ 271
............................................................................................. 271
..............................................................................................272
............................................................................................. 272
............................................................................... 272
.........................................................................................273
FORMat:DEXPort:DSEParator
<Separator>
This command selects the decimal separator for data exported in ASCII format.
Parameters:
<Separator>
POINt | COMMa
COMMa
Uses a comma as decimal separator, e.g.
4,05
.
POINt
Uses a point as decimal separator, e.g.
4.05
.
*RST:
*RST has no effect on the decimal separator.
Default is POINt.
Example:
FORM:DEXP:DSEP POIN
Sets the decimal point as separator.
Manual operation:
See
See
FORMat:DEXPort:FORMat
<FileFormat>
Determines the format of the ASCII file to be imported or exported. Depending on the
external program in which the data file was created or will be evaluated, a comma-sep-
arated list (CSV) or a plain data format (DAT) file may be required.
Parameters:
<FileFormat>
CSV | DAT
*RST:
DAT
Example:
FORM:DEXP:FORM CSV
Manual operation:
See