Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
262
User Manual 1178.8536.02 ─ 01
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks[:IMMediate]
<Peaks>
This command initiates a peak search.
Suffix:
<n>
.
<m>
Parameters:
<Peaks>
This parameter defines the number of peaks to find during the
search.
Note that the actual number of peaks found during the search
also depends on the peak excursion you have set with
CALCulate<n>:MARKer<m>:PEXCursion
Range:
1 to 200
Example:
CALC:MARK:PEXC 5
Defines a peak excursion of 5 dB, i.e. peaks must be at least 5
dB apart to be detected as a peak.
CALC:MARK:FUNC:FPE 10
Initiates a search for 10 peaks on the current trace.
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:LIST:SIZE
<MaxNoPeaks>
This command defines the maximum number of peaks that the R&S
FSV/A looks for
during a peak search.
Suffix:
<n>
.
<m>
Parameters:
<MaxNoPeaks>
Maximum number of peaks to be determined.
Range:
1 to 200
*RST:
50
Example:
CALC:MARK:FUNC:FPE:LIST:SIZE 10
The marker peak list will contain a maximum of 10 peaks.
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:SORT
<SortMode>
This command selects the order in which the results of a peak search are returned.
Suffix:
<n>
.
<m>
Parameters:
<SortMode>
X
Sorts the peaks according to increasing position on the x-axis.
I/Q Analysis