Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
150
User Manual 1178.8536.02 ─ 01
*RST:
1
Example:
SOUR:EXT:FREQ:COUP ON
Manual operation:
See
SOURce<si>:EXTernal<gen>:FREQuency[:FACTor]:DENominator
<Value>
This command defines the denominator of the factor with which the analyzer frequency
is multiplied in order to obtain the transmit frequency of the selected generator.
Select the multiplication factor such that the frequency range of the generator is not
exceeded if the following formula is applied to the start and stop frequency of the ana-
lyzer:
Offset
Analyzer
Generator
F
ator
Deno
Numerator
F
F
min
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<Value>
<numeric value>
*RST:
1
Example:
//Define multiplication factor of 4/3; the transmit frequency of the
generator is 4/3 times the analyzer frequency
SOUR:EXT:FREQ:NUM 4
SOUR:EXT:FREQ:DEN 3
Manual operation:
See
"(Automatic) Source Frequency (Numerator/Denominator/
SOURce<si>:EXTernal<gen>:FREQuency[:FACTor]:NUMerator
<Value>
This command defines the numerator of the factor with which the analyzer frequency is
multiplied in order to obtain the transmit frequency of the selected generator.
Select the multiplication factor such that the frequency range of the generator is not
exceeded if the following formula is applied to the start and stop frequency of the ana-
lyzer:
Offset
Analyzer
Generator
F
ator
Deno
Numerator
F
F
min
Suffix:
<si>
.
irrelevant
<gen>