Analysis
R&S
®
FSV/A3000 I/Q
Analyzer
112
User Manual 1178.8536.02 ─ 01
Remote command:
CALCulate<n>:MARKer<m>:PEXCursion
Search Limits
The search results can be restricted by limiting the search area or adding search con-
ditions.
Search Limits ( Left / Right ) ← Search Limits
If activated, limit lines are defined and displayed for the search. Only results within the
limited search range are considered.
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt
Search Threshold ← Search Limits
Defines an absolute threshold as an additional condition for the peak search. Only
peaks that exceed the threshold are detected.
Remote command:
Use Zoom Limits ← Search Limits
If activated, the peak search is restricted to the active zoom area defined for a single
zoom.
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits:ZOOM[:STATe]
Deactivating All Search Limits ← Search Limits
Deactivates the search range limits.
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
6.4.2.3
Positioning Functions
Access
: [MKR ->]
The following functions set the currently selected marker to the result of a peak search
or set other characteristic values to the current marker value.
................................................................................................ 113
Center Frequency = Marker Frequency
..................................................................... 113
Reference Level = Marker Level
................................................................................ 113
Marker Usage