Remote Commands for Bluetooth Measurements (K8)
R&S
®
FSV-K8
115
Operating Manual 1176.7610.02 ─ 02
Parameters:
<FilterType>
NORMal
Gaussian filters
CFILter
channel filters
RRC
RRC filters
P5
5 Pole filters
*RST:
NORMal
Example:
BAND:TYPE NORM
Manual operation:
See
"Filter Type (Output Power, TX Spec ACLR, Spurious Emis-
[SENSe:]CORRection:EGAin:INPut[:MAGNitude]
This command makes an external gain known to the analyzer, which will take it into
account during the display of measurement results. With this function the gain of an
antenna or of an external preamplifier can be taken into account for the measurement
values.
The command is available if a tracking generator is installed and has been turned on.
Parameters:
Range:
-200 dB to 200 dB
*RST:
0 dB
Example:
CORR:EGA:INP 10DB
Takes 10 dB external gain into account.
Manual operation:
See
[SENSe:]FREQuency:CENTer
<Frequency>
This command defines the center frequency (frequency domain) or measuring fre-
quency (time domain).
Parameters:
<Frequency>
Range:
0 to fmax
*RST:
fmax/2
Default unit: Hz
f
max
is specified in the data sheet. min span is 10 Hz
Example:
FREQ:CENT 100 MHz
Manual operation:
See
SENSe Subsystem (BLUETOOTH, K8)