Remote Control Commands - R&S
FSV-K70
R&S
®
FSV-K70
267
Operating Manual 1176.7578.02 ─ 09
DISPlay[:WINDow<n>]:ZOOM:AREA
<x1>, <y1>, <x2>, <y2>
This command defines the zoom area.
Before you can define a zoom area, you first have to turn the zoom on.
Suffix:
<n>
.
window; For applications that do not have more than 1 measure-
ment window, the suffix <n> is irrelevant.
Parameters:
<x1>,<y1>,
<x2>,<y2>
Diagram coordinates in % of the complete diagram that define
the zoom area.
The lower left corner is the origin of coordinate system. The
upper right corner is the end point of the system.
Range:
0 to 100
Default unit: PCT
Example:
DISP:ZOOM ON
Activates the zoom mode.
DISP:ZOOM:AREA 5,30,20,100
Enlarges the display of the measurement results in the area
defined by the coordinates (5,30) and (20,100).
4.5 FORMat subsystem
........................................................................................ 267
..............................................................................................268
................................................................................................268
FORMat:DEXPort:DSEParator
<Separator>
This command defines which decimal separator (decimal point or comma) is to be
used for outputting measurement data to the file in ASCII format. Different languages
of evaluation programs (e.g. MS-Excel) can thus be supported.
Parameters:
<Separator>
POINt | COMMA
*RST:
(factory setting is POINt; *RST does not affect set-
ting)
FORMat subsystem
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...