Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
161
Operating Manual 1176.7578.02 ─ 09
Auto/On/Off
Enables or disables pattern searches. If "Auto" is selected, pattern search is enabled
automatically if "Pattern" is selected in the "Signal Description" tab of the "Modulation &
Signal Description" dialog box (see
Remote command:
[SENSe:]DDEMod:SEARch:SYNC:MODE
[SENSe:]DDEMod:SEARch:SYNC:AUTO
Meas only if pattern symbols correct
If enabled, measurement results are only displayed (and averaged) if a valid pattern
has been found. For measurements of signals with patterns that are averaged over
several sweeps, this option should be enabled so that erroneous measurements do not
affect the result of averaging.
Remote command:
[SENSe:]DDEMod:SEARch:SYNC:MODE
Auto Configuration
Configures the pattern search automatically. If enabled, the
setting is not available.
Remote command:
[SENSe:]DDEMod:SEARch:PATTern:CONFigure:AUTO
I/Q Correlation Threshold ← Auto Configuration
The I/Q correlation threshold decides whether a match is accepted or not during a pat-
tern search (see also
Chapter 3.3.5, "Working with Pattern Searches"
the parameter is set to 100%, only I/Q patterns that match totally with the input signal
are found. This is only the case for infinite SNR.
The default value is 90%. As long as the pattern is found, there is no need to change
this parameter. However, if the pattern is very short (approximately < 10 symbols) or if
the signal is highly distorted, tuning this parameter helps the pattern search to suc-
ceed.
Remote command:
[SENSe:]DDEMod:SEARch:SYNC:IQCThreshold
Select Pattern for Search
Determines which of the patterns that are assigned to the current standard is to be
searched for. Only one pattern can be selected at a time. However, to check for sev-
eral patterns in the same captured signal, select the single sweep mode (Statistic
Count = 0 or 1) and change the pattern. The measurement is updated.
Remote command:
[SENSe:]DDEMod:SEARch:SYNC:SELect
Configuring VSA Measurements
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...