In the "External" or "IF Power" trigger mode, a common input signal is used for both
trigger and gate. Therefore, changes to the gate delay will affect the trigger delay (trig-
ger offset) as well.
Remote command:
TRIGger<n>[:SEQuence]:HOLDoff[:TIME]
Trigger Level
Defines the trigger level as a numeric value.
Remote command:
TRIGger<n>[:SEQuence]:LEVel:IFPower
For digital input via the R&S Digital I/Q Interface, R&S
FSV-B17:
TRIGger<n>[:SEQuence]:LEVel:BBPower
Trigger Polarity
Sets the polarity of the trigger source.
The sweep starts after a positive or negative edge of the trigger signal. The default set-
ting is "Pos". The setting applies to all modes with the exception of the "Free Run" and
"Time" mode.
"Pos"
Level triggering: the sweep is stopped by the logic "0" signal and
restarted by the logical "1" signal after the gate delay time has
elapsed.
"Neg"
Edge triggering: the sweep is continued on a "0" to "1" transition for
the gate length duration after the gate delay time has elapsed.
Remote command:
Trigger Hysteresis
Defines the value for the trigger hysteresis for "IF power" or "RF Power" trigger sour-
ces. The hysteresis in dB is the value the input signal must stay below the power trig-
ger level in order to allow a trigger to start the measurement. The range of the value is
between 3 dB and 50 dB with a step width of 1 dB.
Remote command:
TRIGger<n>[:SEQuence]:IFPower:HYSTeresis
Trigger Holdoff
Defines the value for the trigger holdoff. The holdoff value in s is the time which must
pass before triggering, in case another trigger event happens.
This softkey is only available if "IFPower", "RF Power" or "BBPower" is the selected
trigger source.
Remote command:
TRIGger<n>[:SEQuence]:IFPower:HOLDoff
For digital input via the R&S Digital I/Q Interface, R&S
FSV-B17:
TRIGger<n>[:SEQuence]:BBPower:HOLDoff
Configuring VSA Measurements
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...