Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
101
Operating Manual 1176.7561.02 ─ 06
The temporary markers T1 and T2 are positioned n dB below the active reference
marker. The frequency and time position of these markers can be queried with
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:FREQuency?
and
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:TIME?
. The bandwidth between
the markers can be queried with
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:
.
Suffix:
<n>
.
Selects the measurement window.
<m>
irrelevant
Parameters:
<Distance>
Distance of the temporary markers to the reference marker in
dB.
*RST:
6dB
Example:
CALC:MARK:FUNC:NDBD 3dB
Sets the level spacing to 3 dB.
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:FREQuency?
This command queries the position of the n dB down markers on the x-axis when mea-
suring in the frequency domain.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m>
irrelevant
Return values:
<Frequency>
<frequency 1>
absolute frequency of the n dB marker to the left of the reference
marker in Hz
<frequency 2>
absolute frequency of the n dB marker to the right of the refer-
ence marker in Hz
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:NDBD ON
Switches on the n dB down function.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:NDBD:FREQ?
Outputs the frequencies of the temporary markers.
Usage:
Query only
Remote Commands of the Analog Demodulation (R&S