Remote Commands
R&S
®
FSMR3000
534
User Manual 1179.0116.02 ─ 02
The search includes all frames. It does not change the horizontal position of the
marker.
If the marker hasn't been active yet, the command first looks for the peak level in the
whole spectrogram and moves the marker vertically to the minimum level.
Suffix:
<n>
.
<m>
12.6.4.7
Fixed Reference Marker Settings
The following commands configure a fixed reference marker.
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:MAXimum[:PEAK]
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:X
............................................. 534
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y
............................................. 535
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y:OFFSet
................................. 535
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed[:STATe]
............................................... 535
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:MAXimum[:PEAK]
This command moves the fixed reference marker to the peak power.
Suffix:
<n>
.
<m>
Example:
CALC:DELT:FUNC:FIX:RPO:MAX
Sets the reference point level for delta markers to the peak of
the selected trace.
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:X
<RefPoint>
This command defines the horizontal position of the fixed delta marker reference point.
The coordinates of the reference may be anywhere in the diagram.
Suffix:
<n>
.
<m>
Parameters:
<RefPoint>
Numeric value that defines the horizontal position of the refer-
ence.
For frequency domain measurements, it is a frequency in Hz.
For time domain measurements, it is a point in time in s.
*RST:
Fixed Reference: OFF
Default unit: HZ
Example:
CALC:DELT:FUNC:FIX:RPO:X 128 MHz
Sets the frequency reference to 128 MHz.
Analyzing Measurements (Basics)