Remote Commands
R&S
®
FSMR3000
453
User Manual 1179.0116.02 ─ 02
[SENSe:]ADEMod:AF:STOP
<Frequency>
This command sets the stop frequency for AF spectrum result display.
Parameters:
<Frequency>
*RST:
9 MHz
Default unit: HZ
Example:
ADEM:AF:STAR 0 kHz
Sets the AF start frequency to 0 kHz
ADEM:AF:STOP 500 kHz
Sets the AF stop frequency to 500 kHz
Manual operation:
See
RF evaluation
These settings are only available for RF evaluation, both in time and frequency
domain.
Useful commands described elsewhere
●
●
Specific commands:
[SENSe:]ADEMod:SPECtrum:SPAN:ZOOM
......................................................................453
[SENSe:]ADEMod:SPECtrum:SPAN[:MAXimum]
...............................................................453
[SENSe:]ADEMod:SPECtrum:SPAN:ZOOM
<Span>
This command sets the span (around the center frequency) for RF spectrum result dis-
play.
The span is limited to the demodulation bandwidth (see
Parameters:
<Span>
*RST:
5 MHz
Default unit: HZ
Example:
ADEM:SPEC:SPAN:ZOOM 200 kHz
Sets the rF span to 200 kHz
[SENSe:]ADEMod:SPECtrum:SPAN[:MAXimum]
<FreqRange>
Sets the DBW to the specified value and the span (around the center frequency) of the
RF data to be evaluated to its new maximum (the demodulation bandwidth).
Parameters:
<FreqRange>
*RST:
5 MHz
Default unit: Hz
Manual operation:
See
Configuring the Measurement