POSitive
Triggers when the signal rises to the trigger level (rising edge).
NEGative
Triggers when the signal drops to the trigger level (falling edge).
*RST:
POSitive
Example:
TRIG:SLOP NEG
Manual operation:
See
TRIGger[:SEQuence]:SOURce
<Source>
This command selects the trigger source.
Note on external triggers:
If a measurement is configured to wait for an external trigger signal in a remote control
program, remote control is blocked until the trigger is received and the program can
continue. Make sure this situation is avoided in your remote control programs.
"Incompleted sequential commands - blocked remote
Parameters:
<Source>
IMMediate
Free Run
EXT | EXT2
Trigger signal from one of the "Trigger Input/Output" connectors.
Note: Connector must be configured for "Input".
*RST:
IMMediate
Example:
TRIG:SOUR EXT
Selects the external trigger input as source of the trigger signal
Manual operation:
See
"Using the power sensor as an external trigger"
on page 116
See
See
See
See
See
See
12.5.6.2
Configuring the Trigger Output
The following commands are required to send the trigger signal to one of the variable
"TRIGGER INPUT/OUTPUT" connectors on the R&S
FSMR3000.
Configuring the Measurement