Network Operation and Remote Control
R&S
®
FSMR3000
327
User Manual 1179.0116.02 ─ 02
Bit No.
Meaning
3
Device-dependent Error
This bit is set if a device-dependent error occurs. An error message with a number between
-300 and -399 or a positive error number, which denotes the error in greater detail, is entered
into the error queue.
4
Execution Error
This bit is set if a received command is syntactically correct but cannot be performed for other
reasons. An error message with a number between -200 and -300, which denotes the error in
greater detail, is entered into the error queue.
5
Command Error
This bit is set if a command is received, which is undefined or syntactically incorrect. An error
message with a number between -100 and -200, which denotes the error in greater detail, is
entered into the error queue.
6
User Request
This bit is set when the instrument is switched over to manual control.
7
Power On (supply voltage on)
This bit is set on switching on the instrument.
11.2.2.4
STATus:OPERation Register
The
STATus:OPERation
register contains information on current activities of the
R&S
FSMR3000. It also contains information on activities that have been executed
since the last read out.
Table 11-5: Meaning of the bits used in the STATus:OPERation register
Bit No.
Meaning
0
CALibrating
This bit is set as long as the instrument is performing a self-alignment.
1-2
Not used
3
SWEeping
Sweep is being performed in base unit (applications are not considered); identical to bit 4
In applications, this bit is not used.
4
MEASuring
Measurement is being performed in base unit (applications are not considered); identical to bit 3
In applications, this bit is not used.
5
Waiting for TRIgger
Instrument is ready to trigger and waiting for trigger signal.
6-7
Not used
8
HardCOPy
in progress
This bit is set while the instrument is printing a hardcopy.
9
not used
Status Reporting System