background image

R&S

©

FPC-Z10 Teaching Kit 

Getting Started 1178.8436.02 – 0

4

Power supply 

The teaching kit has two USB interfaces to supply it with power, a micro USB interface and a 
mini USB interface. You can use both interfaces to supply the teaching kit with power. 

A USB cable for the micro USB interface is included in the delivery. If you want to use the mini 
USB interface for power supply, you have to use a corresponding cable.  

Select the USB interface you want to use as a power supply with the jumper 
(5). 
Position 1:  Selects the micro USB interface as the power supply. 
Position 2:  Select the mini USB interface as the power supply. 
The ratings of the interfaces depend on the application. 

● 

For measurements with an inactive baseband circuitry, the rating is 5 V DC at 450 mA.

Use the USB cable delivered with the teaching kit to connect it to another USB interface. 
You can use the USB interfaces on the R&S

®

FPC as the power source, for example.

● 

For measurements with an active baseband circuitry, the rating is 5 V DC at 750 mA. 

Using the baseband connectors requires an external power supply with a rating of 5 V DC at 
1 A. If an external power supply is required, the LED above the mini USB interface is
illuminated. 

Baseband state 

If you want to use the I/Q modulator or demodulator, you have to activate the baseband signal 
path on the DIP switch (8). 

Switch 7 controls the baseband input state. Move it into the "on" position if 
you want to use the baseband connectors. If you turn on the baseband, 
you have to to use an external power supply as described under "Power 
supply". 
Note that switch 1 has to be always "On". The other switches on this DIP switch are not 
supported. 

Содержание FPC-Z10 Teaching Kit

Страница 1: ...R S FPC Z10 Teaching Kit Getting Started 1178843602 Version 04 Getting Started...

Страница 2: ...164 Email info rohde schwarz com Internet www rohde schwarz com Subject to change Data without tolerance limits is not binding R S is a registered trademark of Rohde Schwarz GmbH Co KG Trade names are...

Страница 3: ...ies to IEC UL EN 60950 1 or IEC UL EN 62368 1 Make sure that the RF signals you apply have a maximum level of 10 dBm and a voltage of 0 V DC Connected equipment must also comply to IEC UL EN 60950 1 o...

Страница 4: ...al routing 4 Jumper for IF PLL power supply selection 5 Power supply selection 6 Calibration standard selection 7 Attenuator for up downconverter 8 Baseband state Connectors A Power supply micro USB B...

Страница 5: ...itry the rating is 5 V DC at 450 mA Use the USB cable delivered with the teaching kit to connect it to another USB interface You can use the USB interfaces on the R S FPC as the power source for examp...

Страница 6: ...apply a signal make sure not to overload the inputs and outputs max 0 dBm You can bypass or close any of the inputs and outputs by setting the corresponding jumpers Here s an example for the downconve...

Страница 7: ...ency with the jumper 1 Position 1 LO frequency of 233 5 MHz Position 2 LO frequency of 636 5 MHz Select this frequency if you are using the internal bandpass filter Synthesizer configuration The synth...

Страница 8: ...upply You can measure the noise of the PLL power supply by measuring the signal on the X105 output The switching converter produces more RF noise than the linear converter Service and maintenance The...

Отзывы: