Operating Manual 1302.6163.12 - 03
4.329
R&S ESU
Instrument Functions
Tracking Generator – Option R&S FSU-B9
4.10.5
Frequency-Converting Measurements
For frequency-converting measurements (e.g. on converter units) the tracking gen-
erator is able to set a constant frequency offset between the output frequency of the
tracking generator and the receive frequency of the R&S ESU.
Up to an output frequency of 200 MHz the measurement can be carried out in both
inverted and normal positions.
Fig. 4.27 Test setup for frequency converting measurements
FREQUENCY
OFFSET
The
FREQUENCY OFFSET
softkey activates the input of the frequency offset
between the output signal of the tracking generator and the input frequency of the
R&S ESU. Possible offsets are in a range of ±200 MHz in 0.1 Hz steps.
The default setting is 0 Hz. Offsets <> 0 Hz are marked with the enhancement label
FRQ
.
If a positive frequency offset is entered, the tracking generator generates an output
signal above the receive frequency of the R&S ESU. In case of a negative frequency
offset it generates a signal below the receive frequency of the R&S ESU. The output
frequency of the tracking generator is calculated as follows:
Tracking generator frequency = receive fre frequency offset.
Remote command:
SOUR:FREQ:OFFS 50MHz
Содержание ESU Series
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...