Remote Control
R&S
®
ESR
796
User Manual 1175.7068.02 ─ 12
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
SWE:EGAT:TRAC1:STAT1 ON
Activates tracing for range 1 of trace 1.
Manual operation:
See
[SENSe:]SWEep:EGATe:TRACe<k>: STOP<range>
<Value>
This command defines the stopping point for the range to be traced using gated trig-
gering
Suffix:
<k>
.
1...6
trace
<range>
1...3
range
Parameters:
<Value>
<numeric value>
*RST:
1 μs
Example:
SWE:EGAT:TRAC1:STOP1 5ms
Sets the stopping point for range 1 on trace 1 at 5 ms.
Manual operation:
See
Scaling the Diagram
The following commands configure the diagram for statistical measurements.
CALCulate<n>:STATistics:PRESet
...................................................................................796
CALCulate<n>:STATistics:SCALe:AUTO ONCE
................................................................ 797
CALCulate<n>:STATistics:SCALe:X:RANGe
.....................................................................797
CALCulate<n>:STATistics:SCALe:X:RLEVel
..................................................................... 798
CALCulate<n>:STATistics:SCALe:Y:LOWer
...................................................................... 798
CALCulate<n>:STATistics:SCALe:Y:UNIT
.........................................................................798
CALCulate<n>:STATistics:SCALe:Y:UPPer
.......................................................................799
CALCulate<n>:STATistics:PRESet
This command resets the scaling of the X and Y axes in a statistical measurement.
The following values are set:
x-axis ref level:
-20 dBm
x-axis range APD:
100 dB
x-axis range CCDF:
20 dB
Remote Commands in Spectrum Analyzer Mode