Remote Control
R&S
®
ESR
744
User Manual 1175.7068.02 ─ 12
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:LIM:ACP:ALT2 30DB, 30DB
Sets the relative limit value for the power in the lower and upper
second alternate adjacent channel to 30 dB below the channel
power.
CALC:LIM:ACP:ALT2:ABS -35DBM, -35DBM
Sets the absolute limit value for the power in the lower and
upper second alternate adjacent channel to -35 dBm.
CALC:LIM:ACP ON
Switches on globally the limit check for the channel/adjacent
channel measurement.
CALC:LIM:ACP:ALT2:STAT ON
Switches on the check of the relative limit values for the lower
and upper second alternate adjacent channel.
CALC:LIM:ACP:ALT2:ABS:STAT ON
Switches on the check of absolute limit values for the lower and
upper second alternate adjacent channel.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:LIM:ACP:ALT2:RES?
Queries the limit check result in the second alternate adjacent
channels.
CALCulate<n>:LIMit<k>:ACPower:ALTernate<Channel>:RESult?
This command queries the result of the limit check for the alternate adjacent channels
for adjacent channel power measurements.
If the power measurement of the adjacent channel is switched off, the command pro-
duces a query error.
Suffix:
<n>
.
Selects the measurement window.
<k>
irrelevant
<Channel>
1...11
the alternate channel
Return values:
<Result>
The result is returned in the form <result>, <result> where
<result> = PASSED | FAILED and where the first (second)
returned value denotes the lower (upper) alternate adjacent
channel.
Remote Commands in Spectrum Analyzer Mode