Remote Control
R&S
®
ESR
690
User Manual 1175.7068.02 ─ 12
If the power measurement in zero span is active, this command limits the evaluation
range to the trace.
Note
: The function is only available if the search limit for marker and delta marker is
switched on (
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
Suffix:
<n>
.
Selects the measurement window.
<m>
irrelevant
Parameters:
<Limit>
The value range depends on the span or sweep time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
Range:
0 to MAX
*RST:
left diagram border
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:RIGH 20MHz
Sets the right limit of the search range to 20 MHz.
Manual operation:
See
See
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
<State>
This command turns marker search limits on and off.
If the power measurement in zero span is active, this command limits the evaluation
range on the trace.
Suffix:
<n>
.
Selects the measurement window.
<m>
marker
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:MARK:X:SLIM ON
Switches on search limitation.
Manual operation:
See
See
See
DISPlay[:WINDow<n>]:MINFo:STATe
<State>
This command turns the marker info field in the diagram on and off.
Remote Commands in Receiver Mode