Remote Control
R&S
®
ESR
689
User Manual 1175.7068.02 ─ 12
Example:
INIT:CONT OFF
Switches to single sweep mode
CALC:MARK:FUNC:ZOOM 1kHz;*WAI
Activates zooming and waits for its end.
Manual operation:
See
CALCulate<n>:MARKer<m>:SCOupled[:STATe]
<State>
This command couples or decouples the marker frequency to the scan range settings.
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
CALC:MARK:SCO ON
Couples the scan range settings to the marker frequency.
Manual operation:
See
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT
<Limit>
This command sets the left limit of the marker search range.
If the power measurement in zero span is active, this command limits the evaluation
range to the trace.
Note
: The function is only available if the search limit for marker and delta marker is
switched on (see
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
Suffix:
<n>
.
Selects the measurement window.
<m>
irrelevant
Parameters:
<Limit>
The value range depends on the span or sweep time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
Range:
0 to MAX
*RST:
left diagram border
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:LEFT 10MHz
Sets the left limit of the search range to 10 MHz.
Manual operation:
See
See
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHT
<Limit>
This command sets the right limit of the marker search range.
Remote Commands in Receiver Mode