Spectrum Measurements
R&S
®
ESR
326
User Manual 1175.7068.02 ─ 12
Adjust Settings ← Scaling ← CCDF
Adjusts the level settings according to the measured difference between peak and min-
imum power for APD measurement or peak and mean power for CCDF measurement
in order to obtain maximum power resolution. Adjusts the reference level to the current
input signal. For details see also the
softkey.
Remote command:
CALCulate<n>:STATistics:SCALe:AUTO ONCE
Gated Trigger (On/Off) ← CCDF
Activates and deactivates the gating for statistics functions for the ACP and the CCDF
channel. The trigger source is changed to "EXTERN" if this function is switched on.
The gate ranges are defined using the
Remote command:
Gate Ranges ← CCDF
Opens a dialog to configure up to 3 gate ranges for each trace.
For details on configuration, see
"Defining Gated Triggering for APD and CCDF"
Remote command:
SWE:EGAT ON
(see
Switches on the external gate mode.
SWE:EGAT:TRAC1:COMM "SlotA"
(see
Adds a comment to trace 1.
SWE:EGAT:TRAC1:STAT1 ON
(see
[SENSe:]SWEep:EGATe:TRACe<k>[:
Activates tracing for range 1 of trace 1.
SWE:EGAT:TRAC1:STAR1 3ms
(see
Sets the starting point for range 1 on trace 1 at 3 ms.
SWE:EGAT:TRAC1:STop1 5ms
(see
Measurements