Operating the R&S
ENV4200
R&S
®
ENV4200
17
User Manual 1107.2470.12 ─ 04
4
Operating the R&S
ENV4200
Before the R&S
ENV4200 is put into operation, read the
Putting the R&S
ENV4200 into operation
Execute the following steps in the order as described. Additional important information
is provided in separate chapters that are linked within each step. See also
1. Place the R&S
ENV4200 on a non-flammable surface and remove all surrounding
items that could obstruct the ventilation openings as described in
2. Establish a protective ground connection of the R&S
"Connecting to protective ground"
3. Establish a reference ground connection of the R&S
"Connecting to reference ground"
4. Connect the equipment under test to the R&S
ENV4200. See
necting the equipment under test"
5. Connect the test receiver to the R&S
ENV4200. See
6.
DANGER!
Risk of fire. Do not operate the R&S
ENV4200 without the auxiliary volt-
age since this voltage is necessary for operation of the fan system. Failure of the
fan system can lead to an overheating of the R&S
ENV4200, burns if the housing
is touched and in extreme cases cause a fire.
To set up the auxiliary voltage supply:
a) Check if the voltage of the auxiliary AC supply corresponds to the selected AC
supply voltage of the R&S
Chapter 4.5, "Connecting the auxili-
on page 20 for further details.
b) Connect the auxiliary AC voltage to the R&S
ENV4200.
7.
DANGER!
Risk of electric shock. See
"Protective ground connection"
Connect the R&S
ENV4200 to the mains supply. See
Chapter 4.7, "Example: Measurement setup"
4.1
Connecting to protective ground
Connect the R&S
ENV4200 to a ground conductor before you put it into operation. See
"Protective ground connection"
on page 5 for further information.
Connecting to protective ground
Содержание ENV4200
Страница 1: ...R S ENV4200 Four Line V Network User Manual 1107247012 Version 04 7H...
Страница 14: ...Instrument tour R S ENV4200 14 User Manual 1107 2470 12 04 2 3 Block diagram Block diagram...
Страница 29: ...Index R S ENV4200 29 User Manual 1107 2470 12 04 Index C Customer support 27 F Front view 9 R Rear View 11...