LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
629
User Manual 1173.9628.02 ─ 16
Manual operation:
See
TRIGger:LTE:MEAS<i>:MEValuation:DELay
<Delay>
Defines a time delaying the start of the measurement relative to the trigger event. This
setting has no influence on "Free Run" measurements.
Parameters:
<Delay>
Range:
-250E-6 s to 250E-6 s
*RST:
0 s
Default unit: s
Example:
See
Configuring the Trigger System
Firmware/Software:
V1.0.10.1
V2.1.20: default unit changed
Manual operation:
See
TRIGger:LTE:MEAS<i>:MEValuation:TOUT
<TriggerTimeOut>
Selects the maximum time that the R&S
CMW will wait for a trigger event before it
stops the measurement in remote control mode or indicates a trigger timeout in manual
operation mode. This setting has no influence on "Free Run" measurements.
Parameters:
<TriggerTimeOut>
Range:
0.01 s to 167772.15 s
*RST:
0.1 s
Default unit: s
Additional parameters: OFF | ON (disables | enables the time-
out)
Example:
See
Configuring the Trigger System
Firmware/Software:
V1.0.10.1
Manual operation:
See
TRIGger:LTE:MEAS<i>:MEValuation:MGAP
<MinTrigGap>
Sets a minimum time during which the IF signal must be below the trigger threshold
before the trigger is armed so that an IF power trigger event can be generated.
Parameters:
<MinTrigGap>
Range:
0 slots to 20 slots
*RST:
2 slots
Default unit: slots
Example:
See
Configuring the Trigger System
Firmware/Software:
V1.0.10.1
Manual operation:
See
Command Reference