LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
616
User Manual 1173.9628.02 ─ 16
<EUTRA>
OFF | ON
*RST:
ON
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V1.0.15.0
Manual operation:
See
3.5.3.6
Power Measurement Settings
The following commands specify settings for the power measurements.
CONFigure:LTE:MEAS<i>:MEValuation:PDYNamics:TMASk
............................................. 616
CONFigure:LTE:MEAS<i>:MEValuation:PDYNamics:AEOPower:LEADing
.......................... 616
CONFigure:LTE:MEAS<i>:MEValuation:PDYNamics:AEOPower:LAGGing
CONFigure:LTE:MEAS<i>:MEValuation:SCOunt:POWer
................................................... 617
CONFigure:LTE:MEAS<i>:MEValuation:POWer:HDMode
.................................................. 617
CONFigure:LTE:MEAS<i>:MEValuation:PDYNamics:TMASk
<TimeMask>
Selects the time mask for power dynamics measurements.
Parameters:
<TimeMask>
GOO | PPSRs | SBLanking
GOO
: General ON/OFF time mask
PPSRs
: PUCCH/PUSCH transmission before and after an SRS
SBLanking
: SRS blanking time mask
*RST:
GOO
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V3.2.70
Manual operation:
See
CONFigure:LTE:MEAS<i>:MEValuation:PDYNamics:AEOPower:LEADing
<Leading>
Shifts the beginning of the evaluation period for OFF power measurements.
Parameters:
<Leading>
Positive values reduce the evaluation period (starts later).
Negative values increase the evaluation period (starts earlier).
Range:
-1000 Ts to 1000 Ts
*RST:
0 Ts
Default unit: Ts
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V3.2.70
Manual operation:
See