CDMA2000 Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
796
User Manual 1173.9511.02 ─ 11
Parameters:
<Repetition>
SINGleshot | CONTinuous
SINGleshot:
Single-shot measurement
CONTinuous:
Continuous measurement
*RST:
SING
Example:
See
Multi Evaluation: Single-Shot and Continuous Measure-
Firmware/Software:
V1.0.2.7
Manual operation:
See
CONFigure:CDMA:MEAS<i>:MEValuation:SCONdition
<StopCondition>
Qualifies whether the measurement is stopped after a failed limit check or continued.
OLFail
means that the measurement is stopped (
STOP:...MEAS<i>...
) and rea-
ches the RDY state as soon as one of the results exceeds the limits.
Parameters:
<StopCondition>
NONE | SLFail
NONE:
Continue measurement irrespective of the limit check
SLFail:
Stop measurement on limit failure
*RST:
NONE
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Firmware/Software:
V1.0.2.7
Manual operation:
See
CONFigure:CDMA:MEAS<i>:MEValuation:SCOunt:MODulation
<SCountMod>
Specifies the statistic count of the measurement. The statistic count is equal to the
number of measurement intervals per single shot. Use
CONFigure:...:MEAS<i>:...:REPetition SINGleshot | CONTinuous
to
select either single-shot or continuous measurements.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
Parameters:
<SCountMod>
Number of measurement intervals.
Range:
1 to 1000
*RST:
1
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Firmware/Software:
V1.0.2.7,
V2.0.10: Change *RST value to 1.
Manual operation:
See
"Statistic Count Modulation / Spectrum"
Command Reference