WCDMA Multi Evaluation Measurement
R&S
®
CMW-KG4xx/-KM4xx/-KS4xx
677
User Manual 1173.9657.02 ─ 11
The results of the current, average, maximum and standard deviation traces can be
retrieved. The standard deviation trace cannot be displayed at the GUI.
See also
chapter 3.2.6.2, "Detailed Views: Modulation, CDP and CDE"
Suffix:
<c>
.
1..2
Selects the carrier to be queried - only relevant for dual carrier
HSUPA
Return values:
<Reliability>
<EVM_1> ...
<EVM_n>
Peak EVM trace results, one result per measured slot or half-slot
Range:
0 % to 100 % (SDEViation: 0 % to 50 %)
Default unit: %
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V1.0.10.1
V3.2.60: command renamed (
CARRier<c>
added)
Options:
R&S CMW-KM405 for dual carrier HSUPA
For additional information concerning syntax elements and returned values refer to
Conventions and General Information
.
FETCh:WCDMa:MEAS<i>:MEValuation:TRACe:EVMagnitude:CHIP:CURRent?
FETCh:WCDMa:MEAS<i>:MEValuation:TRACe:EVMagnitude:CHIP:AVERage?
FETCh:WCDMa:MEAS<i>:MEValuation:TRACe:EVMagnitude:CHIP:MAXimum?
READ:WCDMa:MEAS<i>:MEValuation:TRACe:EVMagnitude:CHIP:CURRent?
READ:WCDMa:MEAS<i>:MEValuation:TRACe:EVMagnitude:CHIP:AVERage?
READ:WCDMa:MEAS<i>:MEValuation:TRACe:EVMagnitude:CHIP:MAXimum?
Returns the values of the RMS EVM vs. chip traces, measured in the preselected slot
(see
on page 645). One value
per chip is returned. The results of the current, average and maximum traces can be
retrieved.
See also
chapter 3.2.6.2, "Detailed Views: Modulation, CDP and CDE"
Return values:
<Reliability>
<EVM1> ...
<EVM2560>
Range:
0 % to 100 %
Default unit: %
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V1.0.4.11
For additional information concerning syntax elements and returned values refer to
Conventions and General Information
.
Command Reference