WCDMA Multi Evaluation Measurement
R&S
®
CMW-KG4xx/-KM4xx/-KS4xx
662
User Manual 1173.9657.02 ─ 11
<Upper>
Range:
0 deg to 90 deg
*RST:
66 deg
Default unit: deg
<Dynamic>
Range:
0 deg to 90 deg
*RST:
36 deg
Default unit: deg
Example:
See
Firmware/Software:
V1.0.4.11
CONFigure:WCDMa:MEAS<i>:MEValuation:LIMit:PHSDpcch
<Enable>,
<MeasurePointA>, <MeasurePointB>, <Dynamic>
Defines a dynamic limit for the phase discontinuity determined by half-slot measure-
ments (signals with HS-DPCCH). The limit is checked at point A and point B. As the
phase discontinuity is measured at half-slot boundaries (x.5, not x.0) point A and point
B should be set to half-slot positions.
Parameters:
<Enable>
OFF | ON
Disables | enables the limit check
*RST:
ON
<MeasurePointA>
Range:
0.5 slots to 119.5 slots
Increment: 0.5 slots
*RST:
0.5 slots
Default unit: slots
<MeasurePointB>
Range:
0.5 slots to 119.5 slots
Increment: 0.5 slots
*RST:
10.5 slots
Default unit: slots
<Dynamic>
Range:
0 deg to 90 deg
*RST:
36 deg
Default unit: deg
Example:
See
Firmware/Software:
V1.0.4.11
CONFigure:WCDMa:MEAS<i>:MEValuation:LIMit:IQOFfset
<IQoffset>
Defines an upper limit for the I/Q origin offset.
Parameters:
<IQoffset>
Range:
-80 dB to 0 dB
*RST:
-25 dB, OFF
Default unit: dB
Additional parameters: OFF | ON (disables the limit check | ena-
bles the limit check using the previous/default limit values)
Command Reference